Tang Jia, Wang Li-qiang, Yuan Bo, et al. Design and calibration of a new high-definition three-dimensional laparoscopic system*[J]. Frontiers of Information Technology & Electronic Engineering, 2015, 16(1): 79-84.
Tang Jia, Wang Li-qiang, Yuan Bo, et al. Design and calibration of a new high-definition three-dimensional laparoscopic system*[J]. Frontiers of Information Technology & Electronic Engineering, 2015, 16(1): 79-84. DOI: 10.1631/FITEE.1400149.
目的
2
为简化三维腹腔镜结构,同时保证其成像和3D视频质量,研发一种新型高清三维腹腔镜系统。
创新
2
采用双光学通道,使用中继透镜延长图像光学传输距离,在中继透镜末端用两个相机接收内窥图像,简化腹腔镜结构,并保证了成像质量。
方法
2
首先,根据图像传感器参数和视场需求大小确定内窥镜光学系统的有效焦距,在此基础上设计包括中继透镜在内的双通道光学镜头(图1)。其次,设计CCU用于采集和处理双通道高清视频图像,并通过HDMI输出将它们同时显示在3D显示器上,由此构建了高清三维腹腔镜系统(图4)。最后,针对双目内窥镜系统采用传统针孔模型计算距离存在较大误差的问题,通过理论分析提出一种新的距离计算方法,并在0-150 mm范围内的距离测量实验中验证该算法的有效性(图3)。
结论
2
新型三维腹腔镜在保证成像质量的基础上简化了系统结构;采用提出的双目系统距离计算方法可以有效提高双目内窥系统的测距精度。
We present a high-definition (HD) 3D laparoscopic system including a dual channel optical system
two cameras
a camera control unit (CCU)
and an HD 3D monitor. This laparoscopic system is capable of outputting dual high-definition videos and providing vivid 3D images. A modified pinhole camera model is used for camera calibration and a new method of depth measurement to improve precision. The average error of depth measurement measured by experiment (about 1.13 mm) was small in proportion to the large range in distance of the system (10–150 mm). The new method is applicable to any calibrated binocular vision system.
JP Barreto , J Roquette , P Sturm , 等 . Automatic camera calibration applied to medical endoscopy . 2009 . Proc. 20th British Machine Vision Conf . 1 - 10 . DOI: 10.5244/C.23.52 http://doi.org/10.5244/C.23.52 .
JY Bouguet . Camera Calibration Toolbox for Matlab . 2013 . Available from http://www.vision.caltech.edu/bouguetj/calib_doc/ [Accessed on July 30, 2014] . .
P Edgcumbe , C Nguan , R Rohling . Calibration and stereo tracking of a laparoscopic ultrasound transducer for augmented reality in surgery . 2013 . Proc. 6th Int. Workshop on Augmented Reality Environments for Medical Imaging and 8th Int. Workshop on Computer-Assisted Interventions . 258 - 267 . DOI: 10.1007/978-3-642-40843-4_28 http://doi.org/10.1007/978-3-642-40843-4_28 .
H Ellis . The Hopkins rod-lens system . J Perioper Pract , 2007 . 17 ( 6 ): 272 - 274 . .
C Feng , JW Rozenblit , AJ Hamilton . A computerized assessment to compare the impact of standard, stereoscopic, and high-definition laparoscopic monitor displays on surgical technique . Surg Endosc , 2010 . 24 ( 11 ): 2743 - 2748 . DOI: 10.1007/s00464-010-1038-6 http://doi.org/10.1007/s00464-010-1038-6 .
M Field , D Clarke , S Strup , 等 . Stereo endoscopy as a 3-D measurement tool . 2009 . Proc. Annual Int. Conf. of the IEEE Engineering in Medicine and Biology Society . 5748 - 5751 . DOI: 10.1109/IEMBS.2009.5332606 http://doi.org/10.1109/IEMBS.2009.5332606 .
J Heikkila , O Silven . A four-step camera calibration procedure with implicit image correction . 1997 . Proc. IEEE Computer Society Conf. on Computer Vision and Pattern Recognition . 1106 - 1112 . DOI: 10.1109/CVPR.1997.609468 http://doi.org/10.1109/CVPR.1997.609468 .
P Honeck , G Wendt-Nordahl , J Rassweiler , 等 . Three-dimensional laparoscopic imaging improves surgical performance on standardized ex-vivo laparoscopic tasks . J Endourol , 2012 . 26 ( 8 ): 1085 - 1088 . DOI: 10.1089/end.2011.0670 http://doi.org/10.1089/end.2011.0670 .
K Keller , A State . A single-imager stereoscopic endoscope . 2011 . SPIE . 79641Z.1 - 79641Z6 . DOI: 10.1117/12.873011 http://doi.org/10.1117/12.873011 .
SH Kong , BM Oh , H Yoon , 等 . Comparison of two- and three-dimensional camera systems in laparoscopic performance: a novel 3D system with one camera . Surg Endosc , 2009 . 24 ( 5 ): 1132 - 1143 . DOI: 10.1007/s00464-009-0740-8 http://doi.org/10.1007/s00464-009-0740-8 .
R Melo , JP Barreto , G Falcao . A new solution for camera calibration and real-time image distortion correction in medical endoscopy—initial technical evaluation . IEEE Trans Biomed Eng , 2012 . 59 ( 3 ): 634 - 644 . DOI: 10.1109/TBME.2011.2177268 http://doi.org/10.1109/TBME.2011.2177268 .
C Wengert , M Reeff , PC Cattin , 等 . Fully automatic endoscope calibration for intraoperative use . 2006 . Proc. des Workshop on Algorithmen Systeme Anwendungen . 419 - 423 . DOI: 10.1007/3-540-32137-3_85 http://doi.org/10.1007/3-540-32137-3_85 .
Z Zhang . A flexible new technique for camera calibration . IEEE Trans Patt Anal Mach Intell , 2000 . 22 ( 11 ): 1330 - 1334 . DOI: 10.1109/34.888718 http://doi.org/10.1109/34.888718 .